cbcvebase.
CVE-2015-3952
published 2019-03-25

CVE-2015-3952: Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq…

PriorityP339high7.5CVSS 3.0
AVNACLPRNUINSUCHINAN
EPSS
1.01%
58.7th percentile
Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

Affected

6 ranges
VendorProductVersion rangeFixed in
hospiraplum_a+3_infusion_system<= 13.6
hospiraplum_a+_infusion_system<= 13.4
hospirasymbiq_infusion_system<= 3.13
pifzerplum_a_+3_infusion_system_firmware<= 13.6
pifzerplum_a_+_infusion_system_firmware<= 13.4
pifzersymbiq_infusion_system_firmware<= 3.13

CVSS provenance

nvdv3.07.5HIGHCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
nvdv2.05.0MEDIUMAV:N/AC:L/Au:N/C:P/I:N/A:N