cbcvebase.
CVE-2018-4845
published 2018-06-26

CVE-2018-4845: A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens…

PriorityP350high8.8CVSS 3.1
AVNACLPRLUINSUCHIHAH
EPSS
1.16%
63.2th percentile
A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions = V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). Remote attackers with either local or remote credentialed access to the "Remote View" feature might be able to elevate their privileges, compromising confidentiality, integrity, and availability of the system. No special skills or user interaction are required to perform this attack. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue.

Affected

3 ranges
VendorProductVersion rangeFixed in
siemensrapidlab_1200_firmware< 3.33.3
siemensrapidpoint_500_firmware<= 2.3
siemensrapidpoint_500_firmware>= 3.0

CVSS provenance

nvdv3.18.8HIGHCVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:H/I:H/A:H
nvdv2.06.5MEDIUMAV:N/AC:L/Au:S/C:P/I:P/A:P
Stop checking back — get the weekly exploitation signal.

Every Monday: what got weaponized or added to CISA KEV in the last seven days — each CVE cross-linked to its PoC, Nuclei template, and detection rule. Free, one email a week, unsubscribe in one click.