CVE-2020-25187
published 2020-12-14CVE-2020-25187: Medtronic MyCareLink Smart 25000 is vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap…
PriorityP260critical9.8CVSS 3.1
AVNACLPRNUINSUCHIHAH
EPSS
3.77%
88.6th percentile
Medtronic MyCareLink Smart 25000 is
vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device
Affected
1 ranges
| Vendor | Product | Version range | Fixed in |
|---|---|---|---|
| medtronic | smart_model_25000_patient_reader | — | — |
CVSS provenance
nvdv3.19.8CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
nvdv2.010.0CRITICALAV:N/AC:L/Au:N/C:C/I:C/A:C
Stop checking back — get the weekly exploitation signal.
Every Monday: what got weaponized or added to CISA KEV in the last seven days — each CVE cross-linked to its PoC, Nuclei template, and detection rule. Free, one email a week, unsubscribe in one click.
No detection rules found.
No public exploits indexed.
No writeups or analysis indexed.
2020-12-14
Published