CVE-2021-33886
published 2021-08-25CVE-2021-33886: An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level…
PriorityP351high8.8CVSS 3.1
AVAACLPRNUINSUCHIHAH
EPSS
0.83%
52.9th percentile
An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
Affected
1 ranges
| Vendor | Product | Version range | Fixed in |
|---|---|---|---|
| bbraun | spacecom2 | < 012u000062 | 012u000062 |
CVSS provenance
nvdv3.18.8HIGHCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
nvdv2.05.8MEDIUMAV:A/AC:L/Au:N/C:P/I:P/A:P
Stop checking back — get the weekly exploitation signal.
Every Monday: what got weaponized or added to CISA KEV in the last seven days — each CVE cross-linked to its PoC, Nuclei template, and detection rule. Free, one email a week, unsubscribe in one click.
No detection rules found.
No public exploits indexed.
No writeups or analysis indexed.
https://www.bbraunusa.com/en.htmhttps://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/https://www.bbraunusa.com/en.htmhttps://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/