cbcvebase.

Bbraun Spacecom2 vulnerabilities

5 known vulnerabilities affecting bbraun/spacecom2.

Total CVEs
5
CISA KEV
0
Public exploits
0
Exploited in wild
0
Severity breakdown
CRITICAL2HIGH3

Vulnerabilities

Page 1 of 1
CVE-2021-33885P2CRITICALCVSS 9.8fixed in 012u0000622021-08-25
CVE-2021-33885 [CRITICAL] CWE-347 CVE-2021-33885: An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U0 An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical dat
nvd
CVE-2021-33884P3CRITICALCVSS 9.1fixed in 012u0000622021-08-25
CVE-2021-33884 [CRITICAL] CWE-434 CVE-2021-33884: An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U An Unrestricted Upload of File with Dangerous Type vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows remote attackers to upload any files to the /tmp directory of the device through the webpage API. This can result in critical files being overwritten.
nvd
CVE-2021-33882P3HIGHCVSS 8.6fixed in 012u0000622021-08-25
CVE-2021-33882 [HIGH] CWE-306 CVE-2021-33882: A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U0000 A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands.
nvd
CVE-2021-33886P3HIGHCVSS 8.8fixed in 012u0000622021-08-25
CVE-2021-33886 [HIGH] CWE-134 CVE-2021-33886: An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a r An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device.
nvd
CVE-2021-33883P3HIGHCVSS 7.5fixed in 012u0000622021-08-25
CVE-2021-33883 [HIGH] CWE-319 CVE-2021-33883: A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U0 A Cleartext Transmission of Sensitive Information vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to obtain sensitive information by snooping on the network traffic. The exposed data includes critical values for a pump's internal configuration.